Sobetracom

Import guide · Tanzania

Importing medical equipment into Tanzania: TMDA registration and import permits

In Tanzania, medical devices, in vitro diagnostics and laboratory equipment come under the Tanzania Medicines and Medical Devices Authority (TMDA). The authority controls two separate things: placing the product on the market, through registration or notification depending on risk, and each individual import, which needs a permit issued per consignment. Both apply, and the import permit is one of the documents the clearing agent must attach to the customs declaration.

For a ministry, hospital or NGO buying equipment in Europe, the timetable therefore has to be worked backwards from the shipping date. This guide sets out the rules published by the TMDA, as of October 2026, along with practical points recorded by the World Food Programme for donations.

Updated on 21 October 2026

Classes A to D: registration or notification

The TMDA classifies devices under the Control of Medical Devices Regulations 2015, using rules aligned with those of the Global Harmonization Task Force. Class A covers low-risk devices such as surgical retractors and tongue depressors; Class B includes hypodermic needles and suction equipment; Class C covers lung ventilators and bone fixation plates; Class D includes heart valves and implantable defibrillators.

Devices in classes A to D are authorised through registration, except selected Class A devices, which only need notification. A notification file includes a description of the product, a 3D presentation, a label using ISO 15223-1 symbols, the manufacturer's ISO 13485 certificate and 2 samples. Registration is valid for 5 years before renewal.

Import permits: apply 6 weeks before shipment

The TMDA asks for permit applications to be lodged at least 6 weeks before shipment, to avoid storage costs at the port of entry or, worse, disposal of goods that arrive without authorisation. The importer's premises must already be registered. Applications are made online through the TMDA trader portal, with processing within 7 days.

The TMDA also sets a shelf-life rule. A device with a shelf life of more than 24 months cannot be imported with less than 60% of it remaining; for a device with a shelf life of 24 months or less, the threshold is 80%. Laboratory consumables and reagents are the most exposed, as a batch made too early can be refused. Devices may only enter through ports of entry authorised by the TMDA.

Special permits for unregistered products

A hospital that needs a device not yet registered in Tanzania still has an option. Section 57 (1) of the Act allows the TMDA to issue a special permit in the public interest. Applicants include hospitals importing for their own use, institutions importing for laboratory or training purposes, exhibitors and importing wholesalers. Conditions are set out in the TMDA's import guidelines for medical devices, and processing is again stated as 7 days.

Donations

The WFP assessment carried out in August 2020 describes the requirements applied to donated health commodities. Both product and supplier had to be registered with the TMDA, which took about 6 months; in an emergency, a registration waiver could be requested on the basis of certificates from the country of origin. The TMDA then approved documents within 24 hours and charged 0.25% of FOB value for the permit. On arrival, its port office inspected the cargo and took samples. These figures date from 2020 and should be reconfirmed.

Frequently asked questions

Must a European manufacturer register its device before an NGO can import it into Tanzania?
As a rule, yes: market access goes through TMDA registration or notification. For hospital or laboratory use, the special permit under section 57 (1) can allow an unregistered product to be imported.
Is CE marking enough?
The TMDA pages consulted do not mention recognition of CE marking. They ask for, among other things, the manufacturer's ISO 13485 certificate and labelling in line with ISO 15223-1.
How does Sobetracom prepare a medical equipment purchase for Tanzania?
From Belgium, Sobetracom checks the device's registration status with the manufacturer, gathers the ISO 13485 certificate and labels, and schedules production so the shelf-life rule is met. TMDA requirements change, so they are reconfirmed with the authority and the buyer's clearing agent before shipping.

Sources

  1. Tanzania Medicines and Medical Devices Authority (TMDA), Importation and Exportation of Medical Devices, Diagnostic Devices and Laboratory Equipment (October 2026)
  2. Tanzania Medicines and Medical Devices Authority (TMDA), Procedure for Marketing Authorization of Medical Devices and In-vitro Diagnostic Devices (October 2026)
  3. Logistics Cluster (WFP), Digital Logistics Capacity Assessment, 1.3.1 Tanzania Medical Customs Information (August 2020)
  4. Tanzania Revenue Authority (TRA), Import Procedures (October 2026)

This page summarises public sources as of the date shown. It does not replace the official text: check it before committing to a purchase, a bid or a shipment.

A project in Africa? Let's talk.

Tell us what you need and where it goes. We come back to you with first leads.