Buying a sterilisation autoclave for a hospital or health centre in Africa
WHO describes pressurised steam as the preferred way to sterilise critical medical devices such as surgical instruments, dressing sets and theatre linen. That makes the autoclave the backbone of any operating theatre, maternity unit or sterile services department, and a breakdown stops clinical work almost at once.
Choosing one is about more than chamber size. The buyer has to settle the type of unit, the cycles needed for the actual loads, how steam will be generated and what water quality is available on site. This page draws on the WHO manual on decontamination and on the European standards for steam sterilisers.
Updated on 20 October 2026
Small or large steriliser
European standards split sterilisers into families. EN 13060:2025 covers small units whose chamber is under 60 litres and cannot take the standard 300 × 300 × 600 mm module. Above that, EN 285:2015+A1:2021 applies to units that hold at least 1 sterilisation module or have a chamber of at least 60 l, the machines found in central sterile services.
WHO notes that table-top sterilisers are the most common type in outpatient, dental and rural clinics, but it discourages their widespread use outside primary and dental care. For a hospital processing theatre trays every day, a row of small units is no substitute for a proper large steriliser.
Matching cycles to loads
EN 13060 distinguishes type B, N and S cycles. As the manufacturer W&H explains them, type B handles every load, including hollow, porous and wrapped items; type N handles only unwrapped solid items; type S handles whatever the steriliser manufacturer has validated. If wrapped instruments or items with lumens are processed, the tender should require a type B cycle.
On temperatures, the WHO manual gives 132 °C to 135 °C for pre-vacuum units, 121 °C or higher for gravity units, and several ranges for pressure-pulse units, including 121 °C to 123 °C. Air removal comes first: without it, steam cannot reach every surface. WHO does not recommend table-top units for instruments with lumens.
Water, steam and site requirements
Water quality is the factor most often overlooked. WHO asks for the supply to be assessed and treated to the manufacturer's specification or, failing that, to ISO 17665, because lime, rust, chlorine and salt deposit on instruments when water is not demineralised, and damage the machine too. A small autoclave runs on distilled or deionised water; a large steriliser needs a steam generator, either built in or connected, and a water treatment unit sized to match.
Before ordering, check the power supply, water inlet and drain, compressed air if the model needs it, ventilation and clearance around the unit for servicing. These constraints shape the choice of model as much as capacity does.
Commissioning, validation and servicing
WHO sets out a staged qualification. For a pre-vacuum autoclave, run 3 cycles with the Bowie-Dick test, then 3 full loaded cycles with biological and chemical controls over 3 consecutive days. Every cycle is then monitored with physical, chemical and biological indicators, and the unit should be serviced at least once a year instead of waiting for it to fail.
The purchase contract should therefore cover operator and technician training, door seals, heating elements, sensors and wear parts, indicators and Bowie-Dick tests for the first period, and full technical documentation. From Belgium, Sobetracom approaches several manufacturers with this specification, compares their declarations of conformity to EN 13060 or EN 285 and their service terms, then coordinates shipping. Our relays in DR Congo, Burundi, Guinea and Nigeria can attend acceptance. Whether the unit must be registered with the national health regulator should be checked before dispatch.
Frequently asked questions
- Does a health centre need a class B autoclave?
- If it sterilises wrapped instruments, linen or hollow instruments, yes: only a type B cycle under EN 13060 covers all of these loads. A type N cycle is enough for unwrapped solid instruments used straight away.
- Can mains water be used in an autoclave?
- The WHO manual calls for water assessed and treated to the manufacturer's requirements or to ISO 17665. Water that is not demineralised leaves lime, rust, chlorine or salt deposits on instruments and in the chamber.
- Which standard applies to a large central steriliser?
- EN 285:2015+A1:2021, which covers units holding at least 1 sterilisation module or with a chamber of at least 60 l. For a small steriliser, the reference is EN 13060:2025.
Sources
- OMS, Decontamination and reprocessing of medical devices for health-care facilities (2016), chapitres 2, 4 et 8 (2016)
- CEN-CENELEC, « Small Steam Sterilizers and Steam Boilers: New Standards for Safer Sterilization » (May 2026)
- Genorma, fiche EN 285:2015+A1:2021 (October 2026)
- W&H (fabricant de stérilisateurs), « Types of sterilizers » (October 2026)
This page summarises public sources as of the date shown. It does not replace the official text: check it before committing to a purchase, a bid or a shipment.