Buying a laboratory analyser for a hospital or health project in Africa

A laboratory analyser is rarely a stand-alone purchase. Behind the instrument come reagents that must be replenished for its whole life, a power supply and sometimes a water feed, regular calibration and a trained technician. Without reagents or servicing the instrument stops, and that is the risk the tender documents should address first.

This guide draws on World Health Organization documents: the sheets in its core medical equipment compendium, the interagency list of priority devices for maternal and child health, and the laboratory quality management handbook, which is based on ISO 15189.

Updated on 21 October 2026

Chemistry or haematology: start from the level of care

A clinical chemistry analyser measures glucose, electrolytes, proteins or drugs in serum, plasma or urine. WHO places it in hospital laboratories, with a version covering blood gases and electrolytes. A haematology analyser counts red cells, white cells and platelets by impedance or light scatter, and WHO's list separates 8 parameter and 18 parameter models depending on the level of the facility.

Sizes vary enormously. According to WHO's compendium, a chemistry analyser weighs 30-700 kg depending on whether it sits on a bench or the floor, whereas a benchtop haematology analyser weighs 15-25 kg and a handheld one 1-5 kg. The right choice depends on daily sample numbers, available skills and what the reference laboratory already covers.

Reagents are the real cost

WHO's handbook puts reagent availability among the first questions to ask, along with whether reagents will be supplied free for a limited period and for how long. The compendium gives a typical reagent shelf life of 1-2 years, and the interagency list states that test kit shelf life must be negotiated as part of the procurement contract.

In the tender, ask for a full list of consumables with part numbers, storage conditions and shelf life. Check too that a dealer in the country or region can supply them and provide service, a selection criterion the handbook names explicitly. An analyser that only takes reagents from one supplier with no regional presence is a poor buy at any price.

Water, power and heat: getting the room ready

A chemistry analyser may need a water supply to rinse its internal parts, on top of mains power and a route for biohazardous waste. WHO points out that temperature and humidity can affect both test kits and analysers, and lists a temperature-controlled work space, reagent refrigeration and electricity or a generator among the requirements. The quality handbook recommends a stable power supply plus backup power for sensitive equipment.

Before ordering, measure doorways and access: the compendium gives approximate dimensions of 500 x 700 x 1,000 mm for a chemistry analyser, and the handbook advises checking door openings and lift access.

Installation, calibration and upkeep

WHO's handbook advises having the manufacturer install the instrument, which often improves the warranty terms, and confirming its responsibilities in writing beforehand. Draw up a checklist of expected performance so it can be verified as soon as the unit is in place, and allow no patient testing until installation is complete, performance has been verified and operators are trained.

Before buying, ask for wiring diagrams, software details, a parts list and the operator's manual, and clarify whether training and the engineer's travel are included. A warranty with a trial period and a maintenance contract should be negotiated at bid stage. After that come a calibration plan and daily, weekly and monthly maintenance over a typical life of 5-7 years for a chemistry analyser and 4-6 years for a haematology analyser.

What Sobetracom does

From Belgium, we turn the laboratory's needs into specifications, approach several manufacturers and compare bids on reagents, regional service and warranty as much as on the instrument, then coordinate shipment of the equipment and its first consumables. In DR Congo, Burundi, Guinea and Nigeria our relays can follow receipt and installation. Reagents are often subject to particular import and transport rules, so have them cleared with the national regulator and your licensed customs broker before dispatch.

Frequently asked questions

Should a laboratory accept an analyser offered by a donor?
WHO's handbook says the laboratory should have a say in donated equipment and should consider declining it if it does not meet its needs, for instance when its reagents cannot be obtained in the country.
What reagent shelf life should we require?
There is no single figure. WHO's compendium gives a typical shelf life of 1-2 years, and the interagency list says shelf life should be negotiated in the contract. Set a minimum remaining shelf life on delivery that fits transit times and the rate of use.

Sources

  1. Organisation mondiale de la santé, Core medical equipment (WHO/HSS/EHT/DIM/11.03), fiches reproduites avec l'autorisation d'ECRI Institute (May 2011)
  2. Organisation mondiale de la santé, Interagency list of priority medical devices for essential interventions for reproductive, maternal, newborn and child health (June 2016)
  3. Organisation mondiale de la santé, CDC et CLSI, Laboratory quality management system: handbook (January 2011)

This page summarises public sources as of the date shown. It does not replace the official text: check it before committing to a purchase, a bid or a shipment.

A project in Africa? Let's talk.

Tell us what you need and where it goes. We come back to you with first leads.