Buying oxygen concentrators for African hospitals: writing the specification

An oxygen concentrator draws in room air, removes the nitrogen and delivers oxygen continuously at the bedside. Because it does not rely on cylinder deliveries, it has become core equipment for health centres and district hospitals. The trade-off is a need for continuous power and regular upkeep.

WHO first published technical specifications for these devices on 9 September 2015, then updated them with UNICEF and again in its 2020 priority devices list. Together, these documents cover what a sound tender needs: performance, environmental conditions, spare parts and the standards manufacturers must prove.

Updated on 20 October 2026

Performance requirements for the tender

The 2020 WHO specification calls for a continuous flow of oxygen above 82 % purity, preferably above 90 %, with a built-in monitor to check it. Maximum flow is chosen to match patient load: 5, 8 or 10 L/min. The flowmeter must go down to 0.5 L/min, which matters on paediatric wards: WHO and UNICEF allow up to 2 L/min per child under 5, and one unit can serve at least 2 children through a flowmeter stand.

Also specify power use of no more than 70 W/L/min, noise below 60 dB(A), an hour meter, alarms for low purity, power failure and high temperature, a weight under 27 kg and castors.

Heat, dust and altitude

The unit must hold its rated purity at 10–40 °C and 15–85 % relative humidity (ideally up to 95 %), from sea level to at least 2000 m. Above that, as on the East African highlands, the manufacturer must state actual performance for its model, so ask in writing before award.

Dust is the enemy of the sieve beds. The WHO-UNICEF specification requires cleanable intake filters and weekly washing of the gross particle filter, more often in very dusty settings, and never with alcohol. Ingress protection of at least IP11, preferably IP21, limits damage from spills. One helpful point for tropical sites: WHO notes that a humidifier may not be needed, since a concentrator supplies oxygen at room temperature.

Power quality causes most failures

WHO and UNICEF report that in many low-resource settings, unstable or low voltage and power surges damage concentrators, sometimes beyond repair. Their minimum recommendation is a voltage stabiliser and a surge suppressor for every installation. The tender should state the local voltage and plug type; the specification asks for safe operation within ± 10 % of rated voltage, and WHO accepts units rated 100–240 V/50–60 Hz with a mains lead of at least 2.5 m. Where outages are frequent, cylinders should be kept as a backup.

Standards, spares and warranty terms

For compliance, WHO lists ISO 80601-2-69 for safety and performance, IEC 60601-1 and IEC 60601-1-2 for electrical safety and electromagnetic compatibility, plus ISO 13485 and ISO 14971 at manufacturer level, along with a free sales certificate from the country of manufacture. Mind the edition: according to the Estonian standards body, ISO 80601-2-69:2020 was withdrawn on 10.04.2026 and replaced by ISO 80601-2-69:2026.

Maintenance should be settled at purchase. The WHO-UNICEF specification asks for filters and parts covering 2 years at 15 hours a day, a warranty of at least 2 years (ideally 5 years) on a device expected to last 7 years, and manuals in English. From Belgium, Sobetracom compares offers from several manufacturers against these criteria, checks certificates and test reports, and bundles filters, sieve beds and stabilisers into the order. Our relays in DR Congo, Burundi, Guinea and Nigeria can follow delivery on site. Registration rules in the destination country still need confirming with the health regulator before ordering.

Frequently asked questions

Is a 5 L/min concentrator enough for a children's ward?
It depends on how many children need oxygen at once. WHO and UNICEF allow up to 2 L/min per child under 5, so a 10 L/min unit with a flowmeter stand serves more patients than a 5 L/min model.
Should every concentrator come with a voltage stabiliser?
WHO recommends one, together with a surge suppressor, as minimum protection wherever mains voltage is unstable. The cost is small compared with replacing a circuit board or compressor.
Which spare parts should be ordered up front?
Internal and external filters, sieve beds, a mains lead, fuses if the unit lacks resettable breakers, and a compressor service kit. The WHO-UNICEF specification sets a minimum of 1 year or 5000 hours of parts, ideally 5 years.

Sources

  1. OMS, Priority medical devices list for the COVID-19 response and associated technical specifications (WHO/2019-nCoV/MedDev/TS/O2T.V2), section 3.3.1.1 (November 2020)
  2. OMS et UNICEF, WHO-UNICEF technical specifications and guidance for oxygen therapy devices (2019), Annex 1 Table A1.2 et chapitre 7 (September 2019)
  3. EVS (organisme estonien de normalisation), fiche ISO 80601-2-69:2020 (October 2026)
  4. OMS, WHO technical specifications for oxygen concentrators (September 2015)

This page summarises public sources as of the date shown. It does not replace the official text: check it before committing to a purchase, a bid or a shipment.