Buying a vaccine refrigerator for Africa: WHO PQS E003 prequalification
The starting point for any vaccine refrigerator is the World Health Organization's PQS (Performance, Quality and Safety) prequalification list, category E003. UNICEF draws on that list for the models it buys under long-term arrangements for immunisation programmes. Where a health centre has no grid connection, or fewer than 8 hours of reliable power a day, WHO points towards solar direct drive (SDD) appliances, which run without a battery.
This guide is built on WHO specification E003/RF05.7, issued on 16 February 2024, UNICEF Supply Division's procurement guidelines and the public data sheet of a prequalified model.
Updated on 20 October 2026
What E003 prequalification already covers
A prequalified model must hold vaccines between +2°C and +8°C at an ambient +43°C and through a +43°C/+25°C day and night cycle. This is the hot zone, which is compulsory for solar appliances. It must also cope with a room that drops to +10°C or below, and no excursion may go above +20°C.
To prevent freezing, the specification grades appliances A, B or C according to how much the user must do to prevent freezing; Grade A needs no intervention at all, wherever a vial sits in the compartment. The thermostat must not be user-adjustable, and every unit carries a built-in data logger with an M2M interface plus a temperature display that can be read with the door shut.
Solar direct drive: assess the site first
UNICEF considers SDD suitable for remote sites, sites off the grid or with fewer than 8 hours of power a day, provided they receive at least 3.5 kWh/m2 of sunlight daily. Energy is stored without batteries using one of 4 approaches: phase change material, ice lining, water lining or an ice bank.
Autonomy is the deciding figure. WHO requires at least 3 days in the hot zone, but a cloudy site may need more, and a unit is only suitable where the site requirement does not exceed its own autonomy. Siting matters as much: panels face south in countries north of the equator and north in those to the south, with no trees casting shade.
Clauses for the tender documents
Ask for the model's PQS number and check its validity date, which every data sheet shows. Then copy the specification's obligations into the contract: a 3-year replacement warranty, a low-maintenance life of at least 10 years, spare parts available for at least 5 years after production ends, 10 spare fuses attached to the unit and a lock supplied with 2 keys. On electrical safety, the manufacturer certifies compliance with IEC 60335-1, IEC 60335-2-24 and IEC 60364-1, and with the electromagnetic compatibility standards IEC 61000-6-1 and IEC 61000-6-3.
PQS data sheets publish measured values, which makes bids comparable. One prequalified combined unit, for instance, lists 57 L of vaccine storage, an autonomy of 82:48 in hours and minutes and R600a refrigerant. Also ask about manual languages and in-country training.
Shipping, receipt and installation
Units must withstand transport and storage between -30°C and +70°C and humidity from 5% to 95%, in export packaging rated at no less than 17kN edge crush resistance. The smallest dimension may not exceed 710 mm, or 830 mm by exception, so that the unit fits through doors and corridors, and WHO recommends that nobody should have to carry more than 25 kg. The combined unit mentioned above ships at 199 kg and 1.33 m3, so plan for enough hands at the receiving end.
UNICEF stresses that these appliances must always travel upright and be left for at least 3 hours before switching on. Allow 1 to 15 weeks from order to delivery at the port of shipment, plus sea transit, and then 4 to 6 weeks between declaring a site ready and the installer's technicians arriving.
How Sobetracom handles this kind of purchase
Working from Belgium, we start from the current PQS list, set aside models whose data sheet has lapsed, and compare manufacturers' bids on autonomy, freeze protection grade, spare parts kit and training terms. We then coordinate packing, shipment and delivery. In DR Congo, Burundi, Guinea and Nigeria our relays can follow acceptance on site. Before the order goes out, have the national immunisation programme and your licensed customs broker confirm the country's own import requirements.
Frequently asked questions
- What does Grade A mean on a vaccine refrigerator?
- It is the highest freeze protection class defined by WHO: the user does not need to do anything to keep vaccines from freezing, wherever they are stored in the compartment. Grades B and C require one or more manufacturer instructions to be followed, such as always using baskets.
- How much autonomy should we require for an off-grid health post?
- At least 3 days, the minimum WHO sets for a solar direct drive refrigerator in the hot zone. Where the rainy season brings long overcast spells, work out the autonomy the site itself needs and compare it with the figure on the model's PQS data sheet.
- Can the unit be switched on as soon as it arrives?
- No. UNICEF asks for it to be carried upright and left for at least 3 hours so that the compressor oil settles. A unit laid on its side in transit can be rendered unusable.
Sources
- Organisation mondiale de la santé, PQS performance specification WHO/PQS/E003/RF05.7 (February 2024)
- UNICEF Supply Division, Cold Chain Support Package, Procurement Guidelines Solar Direct Drive Refrigerators and Freezers (December 2020)
- OMS, Immunization Devices PQS, fiche produit E003-084e (October 2026)
This page summarises public sources as of the date shown. It does not replace the official text: check it before committing to a purchase, a bid or a shipment.